期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
Implementation and uniformity calibration of LED array for photodynamic therapy 被引量:1
1
作者 Xu Wang Jiyuan Xiong +1 位作者 Xiaoming Hu Qin Li 《Journal of Innovative Optical Health Sciences》 SCIE EI CAS 2022年第4期61-69,共9页
Irradiance uniformity is an important parameter for a Photodynamic Therapy(PDT)device.The calibration and verification of a LED array light source based on computer vision technology were implemented and carried out.F... Irradiance uniformity is an important parameter for a Photodynamic Therapy(PDT)device.The calibration and verification of a LED array light source based on computer vision technology were implemented and carried out.First,the correlation coe±cients between the pulse width modulate value and the irradiance were calibrated.Then,the correction of the actual light center and divergence angle were solved by image processing to reduce errors from each LED lens.Finally,uniformity was optimized according to the irradiance formula of the Lambertian source.The lowest coe±cients of variation of irradiance were 4.87%in a 5 cm×12 cm area and 3.55%in a 3 cm×10 cm area within the depth range of 8–12 cm when the expected irradiance was 100 mW/cm^(2).This finding indicated that the light source can achieve a more uniform illumination and provide a better therapeutic effect for the PDT of port-wine stains. 展开更多
关键词 Photodynamic therapy computer vision light source calibration