Rehabilitation of lower limb amputation in developing countries is quite challenging.Though there are basic to highly advanced prostheses available in India,the set-up is still facing difficulties in developing countr...Rehabilitation of lower limb amputation in developing countries is quite challenging.Though there are basic to highly advanced prostheses available in India,the set-up is still facing difficulties in developing countries.Prosthetic management is difficult due to lack of availability of prostheses and reduced affordability among low income populations.In this review we highlighted the lower limb amputation and prosthetic rehabilitation status in India.Currently,India is advancing well in the rehabilitation field,but further studies are required to provide more evidence and recommendation.展开更多
This paper presents the design and control of a pneumatically actuated transtibial prosthesis, which utilizes a pneumatic cylinder-type actuator to power the prosthetic ankle joint to support the user's locomotion. T...This paper presents the design and control of a pneumatically actuated transtibial prosthesis, which utilizes a pneumatic cylinder-type actuator to power the prosthetic ankle joint to support the user's locomotion. The pneumatic actuator has multiple advantages over the traditional electric motor, such as light weight, low cost, and high power-to-weight ratio. The objective of this work is to develop a compact and lightweight transtibial prosthesis, leveraging the multiple advantages provided by this highly competitive actuator. In this paper, the design details of the prosthesis are described, including the determination of performance specifications, the layout of the actuation mechanism, and the calculation of the torque capacity. Through the authors' design calculation, the prosthesis is able to provide sufficient range of motion and torque capacity to support the lo- comotion of a 75 Kg individual. The controller design is also described, including the underlying biomechanical analysis and the formulation of the finite-state impedance controller. The testing results of human subject indicate that the prosthesis is able to generate a natural walking gait and sufficient power output for its amputee user.展开更多
文摘Rehabilitation of lower limb amputation in developing countries is quite challenging.Though there are basic to highly advanced prostheses available in India,the set-up is still facing difficulties in developing countries.Prosthetic management is difficult due to lack of availability of prostheses and reduced affordability among low income populations.In this review we highlighted the lower limb amputation and prosthetic rehabilitation status in India.Currently,India is advancing well in the rehabilitation field,but further studies are required to provide more evidence and recommendation.
基金The authors gratefully acknowledge the support of the National Institutes of Health under Grant No. 5R01HD075493 and the National Science Foundation under Grant No. 1351520.
文摘This paper presents the design and control of a pneumatically actuated transtibial prosthesis, which utilizes a pneumatic cylinder-type actuator to power the prosthetic ankle joint to support the user's locomotion. The pneumatic actuator has multiple advantages over the traditional electric motor, such as light weight, low cost, and high power-to-weight ratio. The objective of this work is to develop a compact and lightweight transtibial prosthesis, leveraging the multiple advantages provided by this highly competitive actuator. In this paper, the design details of the prosthesis are described, including the determination of performance specifications, the layout of the actuation mechanism, and the calculation of the torque capacity. Through the authors' design calculation, the prosthesis is able to provide sufficient range of motion and torque capacity to support the lo- comotion of a 75 Kg individual. The controller design is also described, including the underlying biomechanical analysis and the formulation of the finite-state impedance controller. The testing results of human subject indicate that the prosthesis is able to generate a natural walking gait and sufficient power output for its amputee user.