The direct detection of clenbuterol(CL) in pig liver without any extraction separation at a pyrrole-DNA modified boron-doped diamond(BDD) electrode is reported. The pyrrole-DNA modified BDD electrode has a strong ...The direct detection of clenbuterol(CL) in pig liver without any extraction separation at a pyrrole-DNA modified boron-doped diamond(BDD) electrode is reported. The pyrrole-DNA modified BDD electrode has a strong electrocatalytic effect on the redox reaction of CL. One oxidization and two reduction peaks of CL appear at 340. 2, 299. 8 and 166. 6 mV( versus SCE), respectively. The pyrrole polymer alone cannot electrocatalyze the above reaction at a BDD electrode ; the electrocatalytic effect of a BDD electrode modified with DNA membrane is unsufficient for the analytical detection of CL; the replacement of boron-doped diamond by glass carbon makes the electrocatalytic reaction impossible ; the redox process is pH dependent. The influences of various experimental parameters on the pyrrole-DNA modified BDD electrode were investigated. A sensitive cyclic vohammetric response for CL was obtained in a linear range from 3.4 × 10^-6 to 5 × 10^ -4 mol/L with a detection limit of 8.5 × 10^-7 mol/L. A mean recovery of 102. 7% of CL in the pig liver sample solution and a reproducibility of 3.2% were obtained.展开更多
基金Supported by the National Natural Science Foundation of China(Nos. 20435010, 20375012, 20205005 and 20475014).
文摘The direct detection of clenbuterol(CL) in pig liver without any extraction separation at a pyrrole-DNA modified boron-doped diamond(BDD) electrode is reported. The pyrrole-DNA modified BDD electrode has a strong electrocatalytic effect on the redox reaction of CL. One oxidization and two reduction peaks of CL appear at 340. 2, 299. 8 and 166. 6 mV( versus SCE), respectively. The pyrrole polymer alone cannot electrocatalyze the above reaction at a BDD electrode ; the electrocatalytic effect of a BDD electrode modified with DNA membrane is unsufficient for the analytical detection of CL; the replacement of boron-doped diamond by glass carbon makes the electrocatalytic reaction impossible ; the redox process is pH dependent. The influences of various experimental parameters on the pyrrole-DNA modified BDD electrode were investigated. A sensitive cyclic vohammetric response for CL was obtained in a linear range from 3.4 × 10^-6 to 5 × 10^ -4 mol/L with a detection limit of 8.5 × 10^-7 mol/L. A mean recovery of 102. 7% of CL in the pig liver sample solution and a reproducibility of 3.2% were obtained.