作者 ZU DongLin,HONG LiMing,CAO XueMing & TANG Xin Institute of Heavy Ion Physics,Beijing City Key Laboratory of Medical Physics and Engineering,School of Physics,Peking University,Beijing 100871,China 《Science China(Technological Sciences)》 SCIE EI CAS 2010年第4期886-891,共6页
In this paper the analytical expressions for the magnetic field H and induction B in iron-pole plates generated by MRI gradient coil are given using line-current and the multilayer dielectric plate model with the mirr... In this paper the analytical expressions for the magnetic field H and induction B in iron-pole plates generated by MRI gradient coil are given using line-current and the multilayer dielectric plate model with the mirror-image method.Eddy current emanates from the magnetic flux in the iron-pole plates.In order to fully suppress the eddy current,this magnetic flux should be fully eliminated.The research results indicate the magnetic permeability of the resist-eddy plate must be bigger than that of magnetic pole material,i.e.pure iron,and that the resist-eddy plate should be thick enough to be far away from its magnetic saturation. 展开更多
关键词 MRI SUPPRESS EDDY current GRADIENT coil resist-eddy plate mirror-image method