Adaptive preprocessing algorithms of corneal topography in polar coordinate system 被引量:1
1
作者 郭雁文 《Journal of Central South University》 SCIE EI CAS 2014年第12期4571-4576,共6页
New adaptive preprocessing algorithms based on the polar coordinate system were put forward to get high-precision corneal topography calculation results. Adaptive locating algorithms of concentric circle center were c... New adaptive preprocessing algorithms based on the polar coordinate system were put forward to get high-precision corneal topography calculation results. Adaptive locating algorithms of concentric circle center were created to accurately capture the circle center of original Placido-based image, expand the image into matrix centered around the circle center, and convert the matrix into the polar coordinate system with the circle center as pole. Adaptive image smoothing treatment was followed and the characteristics of useful circles were extracted via horizontal edge detection, based on useful circles presenting approximate horizontal lines while noise signals presenting vertical lines or different angles. Effective combination of different operators of morphology were designed to remedy data loss caused by noise disturbances, get complete image about circle edge detection to satisfy the requests of precise calculation on follow-up parameters. The experimental data show that the algorithms meet the requirements of practical detection with characteristics of less data loss, higher data accuracy and easier availability. 展开更多
关键词 corneal topography Placido disk polar coordinate self-adoption preprocessing algorithms