This paper reported a compact system of capacitively coupled contactless conductivity detection (C4D) based on the square wave excitation voltage for capillary electrophoresis, and it exhibited excellent sensitivity...This paper reported a compact system of capacitively coupled contactless conductivity detection (C4D) based on the square wave excitation voltage for capillary electrophoresis, and it exhibited excellent sensitivity at the optimal frequency of 198 kHz. The feasibility and sensitivity of this detector was demonstrated by simultaneous detection of thirteen ions including alkali, alkaline earth and heavy metal ions. And the detection limits (S/N 3) were in the range of 0.2-1μmol/L for Mn^2+, K^+, Na^+, Mg^2+, Ca^2+, Li^+, Ba^2+, and 7-25 μmol/L for Ni^2+, Cu^2+, Cd^2+, Pb^2+, Co^2+, Zn^2+.展开更多
基金support from the National Natural Science Foundation of China(No.20475018)the Key Program of Guangdong Natural Science Foundation(No.05103552)
文摘This paper reported a compact system of capacitively coupled contactless conductivity detection (C4D) based on the square wave excitation voltage for capillary electrophoresis, and it exhibited excellent sensitivity at the optimal frequency of 198 kHz. The feasibility and sensitivity of this detector was demonstrated by simultaneous detection of thirteen ions including alkali, alkaline earth and heavy metal ions. And the detection limits (S/N 3) were in the range of 0.2-1μmol/L for Mn^2+, K^+, Na^+, Mg^2+, Ca^2+, Li^+, Ba^2+, and 7-25 μmol/L for Ni^2+, Cu^2+, Cd^2+, Pb^2+, Co^2+, Zn^2+.