Design and gamma sensitivity measurement of a novel dual-emitter vacuum Compton detector
1
作者 韩和同 王群书 +2 位作者 夏良斌 管兴胤 张子川 《Chinese Physics B》 SCIE EI CAS CSCD 2009年第11期4777-4780,共4页
A novel dual-emitter vacuum Compton detector (D-VCD) with higher gamma ray detecting efficiency is proposed. The emitters are made of Ta A1 clad metal. The gamma ray sensitivity is studied by Monte Carlo simulation ... A novel dual-emitter vacuum Compton detector (D-VCD) with higher gamma ray detecting efficiency is proposed. The emitters are made of Ta A1 clad metal. The gamma ray sensitivity is studied by Monte Carlo simulation using the MCNP code. A comparison between calculations and results measured by using the 1.25 MeV gamma ray of Co-60 is also performed. Experimental sensitivities for two sample D-VCDs with the same materials and structures are 1.92×10^-20 and 2.02×10^-20·cm^2/MeV separately, which are consistent with the simulation result of 1.98×10^-20·cm^2/MeV and are 4 times higher than that of VCD with a single Fe emitter. According to the simulation results, in a gamma energy range from 0.5 to 3 MeV, the maximum sensitivity variance for the D-VCD is less than 15%, and less than 5% in a range from 1 to 2 MeV in particular. The novel D-VCD is applicable to the detection of intense pulse gamma rays. 展开更多
关键词 vacuum compton detector sensitivity pulsed gamma ray
下载PDF
相对永久密封真空康普顿探测器伽马灵敏度研究 被引量:3
2
作者 韩和同 王群书 +4 位作者 夏良斌 熊华 管兴胤 张子川 张侃 《西安交通大学学报》 EI CAS CSCD 北大核心 2009年第3期70-73,共4页
针对强流脉冲γ射线探测,研制了厚入射窗结构真空康普顿探测器,并在^60Co辐照装置上,准直孔径为40mm的条件下,对2套样品探测器灵敏度进行了测量,得到了1.25MeV的了射线灵敏度,实验值分别为4.65×10^-22C/MeV和4.59×10... 针对强流脉冲γ射线探测,研制了厚入射窗结构真空康普顿探测器,并在^60Co辐照装置上,准直孔径为40mm的条件下,对2套样品探测器灵敏度进行了测量,得到了1.25MeV的了射线灵敏度,实验值分别为4.65×10^-22C/MeV和4.59×10^-22C/MeV.该实验值与蒙特卡罗程序MCNP的计算值(4.70×10^-22C/MeV)相比,误差范围一致,在0.5~3.0MeV能量区间,探测灵敏度的能量响应变化小于20%.实验证明:该探测器比薄窗结构真空康普顿探测器的灵敏度提高了60%,克服了薄窗探测器封装加工和静态真空保持的困难;保存使用期可达2年以上,达到了相对永久密封的目的,可应用于粒子注量率为10^20cm^-2·s^-1以上的强流脉冲γ射线测量. 展开更多
关键词 脉冲 伽马射线 康普顿探测器 灵敏度 蒙特卡罗程序
下载PDF
上一页 1 下一页 到第
使用帮助 返回顶部