Research on Moflo XDP high speed cell sorting techniques and parameter optimization 被引量:1
1
作者 Yufang Sun Chen Zheng Bo Xu 《Journal of Chinese Pharmaceutical Sciences》 CAS CSCD 2020年第5期355-363,共9页
This article aims to discuss strategies to operate Moflo XDP high-speed sorting equipment for cell sorting and optimize the instrument to run in the best operating state.Using fluorescent microspheres to quickly calib... This article aims to discuss strategies to operate Moflo XDP high-speed sorting equipment for cell sorting and optimize the instrument to run in the best operating state.Using fluorescent microspheres to quickly calibrate the position of the main liquid flow and the laser beam,adjust the amplitude and frequency according to different size CytoNozzles to find the best breakoff point,and adjust the four-way side flow to achieve the stable maintaining of the equipment,so as to quickly sort sample at high-speed.The absence of air bubbles in the sheath liquid barrel and the liquid path is the key to achieving a stable sorting-maintaining in the sorting process.The accurate calculation of the breakoff point and drop delay is the key to improve sorting efficiency and cell activity. 展开更多
关键词 Cell sort BUBBLE Maintaining CV optimization