期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
Sound transducer calibration of ambulatory audiometric system utilizing delta learning rule
1
作者 KIM Kyeong-seop SHIN Seung-won +3 位作者 YOON Tae-ho LEE Sang-min LEE Insung RYU Keun ho 《Journal of Central South University》 SCIE EI CAS 2011年第6期2009-2014,共6页
An efficient calibration algorithm for an ambulatory audiometric test system is proposed. This system utilizes a personal digital assistant (PDA) device to generate the correct sound pressure level (SPL) from an audio... An efficient calibration algorithm for an ambulatory audiometric test system is proposed. This system utilizes a personal digital assistant (PDA) device to generate the correct sound pressure level (SPL) from an audiometric transducer such as an earphone. The calibrated sound intensities for an audio-logical examination can be obtained in terms of the sound pressure levels of pure-tonal sinusoidal signals in eight-banded frequency ranges (250, 500, 1 000, 2 000, 3 000, 4 000, 6 000 and 8 000 Hz), and with mapping of the input sound pressure levels by the weight coefficients that are tuned by the delta learning rule. With this scheme, the sound intensities, which evoke eight-banded sound pressure levels by 5 dB steps from a minimum of 25 dB to a maximum of 80 dB, can be generated without volume displacement. Consequently, these sound intensities can be utilized to accurately determine the hearing threshold of a subject in the ambulatory audiometric testing environment. 展开更多
关键词 传感器校准 声音强度 规则系统 三角洲 门诊 听力 个人数字助理 声压级
下载PDF
上一页 1 下一页 到第
使用帮助 返回顶部