Development of a Poly(vinyl chloride)-based Nitrate Ion-selective Electrode
1
作者 HE Meng-meng YANG Hui-zhong 《Chemical Research in Chinese Universities》 SCIE CAS CSCD 2012年第6期957-962,共6页
The authors developed a nitrate ion-selective electrode(ISE) based on poly(vinyl chloride)(PVC) membrane with methyltrioctylammonium nitrate as a carrier and 1-decanol as a plasticizer.The performance of the nit... The authors developed a nitrate ion-selective electrode(ISE) based on poly(vinyl chloride)(PVC) membrane with methyltrioctylammonium nitrate as a carrier and 1-decanol as a plasticizer.The performance of the nitrate-sensitive membranes was optimized by tuning the composition of components.The electrode exhibits a linear response with a Nernstian slope of(52±1.0) mV per decade for the nitrate ion concentration ranging from 5.8×10-5 mol/L to 1.0 mol/L.The electrode can be used to detect a low concentration of nitrate ions down to 3×10-5 mol/L in a pH range of 2.1―11.5 without any compensation.The advantage of the electrode includes simple preparation,short response time and good repeatability.The detection performance of the novel electrode on nitrate ions has been tested for water samples. 展开更多
关键词 Poly(vinyl chloride)(PVC) Methyltrioctylammonium nitrate Nitrate-selective electrode Potentiometric sensor