Development of a novel endoscopic manipulation system: The Endoscopic operation robot 被引量:4
1
作者 Keiichiro Kume Takeshi Kuroki +1 位作者 Takahiro Sugihara Masafumi Shinngai 《World Journal of Gastrointestinal Endoscopy》 CAS 2011年第7期145-150,共6页
AIM: To develop and evaluate the endoscopic operation robot (EOR). The EOR is a robot system designed specifically for remote manipulation of the scope during gastrointestinal endoscopy by a seated endoscopist. METHOD... AIM: To develop and evaluate the endoscopic operation robot (EOR). The EOR is a robot system designed specifically for remote manipulation of the scope during gastrointestinal endoscopy by a seated endoscopist. METHODS: Total colonoscopy examinations using a colonoscopy training model were performed compared conventional insertion by manual manipulation and remote-controlled insertion, using the EOR. The author investigated the time taken for each of the 50 examinations. RESULTS: The median insertion time (in minutes) for each 10 examinations (EOR vs manual manipulation) was 73.70 ± 25.37 vs 3.77 ± 1.34 in the first group, 38.40 ± 6.24 vs 3.40 ± 0.97 in the second group, 27.6 ± 4.01 vs 2.70 ± 0.95 in the third group, 23.8 ± 3.65 vs 3.10 ± 0.88 in the fourth group, and 22.9 ± 5.02 vs 2.60 ± 1.08 in the fifth group.CONCLUSION: The study suggested the possibility of the clinical application of the EOR. 展开更多
关键词 Hepatitis B Hepatocellular cancer Hepatitis B surveillance VACCINATION SCREENING